Skip to main content

Select your region & language

Global

Region

CF-5200 Frequently Asked Questions

Question number FAQ-0371

Phase data becomes zero in the harmonic list.

When the spectrum is displayed on the Y-axis in dB, the phase of the spectrum below the lower limit of the Y-axis scale is forcibly treated as 0 degrees. Therefore, areas where the spectrum (Mag) values are small are cut off (displaying 0), resulting in a reduced number of items in the list. (Please excuse the lack of aesthetic quality in the figure.)

Please use the function in section 6-1-2 of the manual "Change the Y-axis scale arbitrarily" to modify the lower limit of the Y-axis.

Last updated: 2000-04-10