Skip to main content

Select your region & language

Global

Region

CF-5200 Frequently Asked Questions

Question number FAQ-0369

What kind of processing is involved in the tracking settings for 'ORDER PEAK' and 'BAND WIDTH'?

Both 'ORDER PEAK' and 'BAND WIDTH' are functions that aim to accurately track the magnitude of order components. They are effective when the order does not perfectly match the set analysis order.

  • ORDER PEAK
    ORDER PEAK is a function that adopts the largest value among three points, including the specified order and the adjacent orders (the adjacent lines of the FFT analysis results) for the number of times to be tracked.
  • BAND WIDTH
    BAND WIDTH is a function that adopts a partial overall value based on the specified number of times centered around the number of times being tracked, set to the number of lines defined by BAND WIDTH. The setting value "0" turns the function "OFF", and it becomes effective at "3" or more. Additionally, the BAND WIDTH function is also effective with ORDER PEAK OFF. Furthermore, when ORDER PEAK is ON and BAND WIDTH is also ON, the BAND WIDTH function takes priority.

Last updated: 2008-03-17