CF-4500 Frequently Asked Questions
Question number FAQ-0296
To start with an external signal and obtain the average result over 10 seconds:
For external average starting, use the function of the input contact (INPUT RELAY) on the rear panel.
The average over 10 seconds is set using the average time function of the averaging processing function.
■ Setting the average processing
Soft keys
MAIN => ANALYSIS => AVERAGE => Select TIME as the TYPE.
The average processing time will be calculated over the set period.
■ Setting the time
Soft keys
MAIN => ANALYSIS => AVERAGE => SET => Open with IME.
Set the time to 10 seconds in the dialog box.
■INPUT RELAY settings
Register the command for average start.
For registration instructions, please refer to the FAQ "External Relay Settings" (PDF).
Short-circuiting INPUT RELAY 1 and COM will start the averaging process.
Last updated: 2013-12-06