CF-4500 Frequently Asked Questions

Question number FAQ-0293

How to save, restore, or copy the settings of the CF-4500 to another CF-4500.

When upgrading the software of the CF-4500 main unit, please back up the panel condition memory and block memory before proceeding with the upgrade.
Performing the upgrade will reset all settings to their defaults.



Operation Procedure

1. Backup from internal memory to USB memory.

「MAIN」→「MEMORY」→「UTILITY」→「COPY FILES」→「COPY FILES TYPES」→「DATA」、

Turn both PANEL COND to ON.

Return to the hierarchy of "COPY FILES" and execute the on-screen start copy with Yes in "INTERNAL⇒USB".

(Please refer to the section on executing data copying in the reference guide.)

2. - Note of items not stored in the panel condition memory -

The following settings will not be saved in the panel condition memory, so please take note of the settings in advance,
and after the update is complete, please manually reset them.

2.1. Items that are not stored in the panel condition memory but are retained even when the power is turned off.

"MAIN" → "CONDITION" below * Includes RS-232C settings.

"MAIN" → "MEMORY" below ※ START TYPE (device startup) will be NEW.

The settings under "CONDITION" and "MEMORY" will be retained regardless of the power ON/OFF, unless a version upgrade is performed.

2.2. Settings that are not stored in the panel condition memory and are initialized at every startup.

"MAIN" → "OUTPUT" → "PHONE" below
"MAIN" → "ANALYSIS" → "CALC" below
"MAIN" → "TRACKING" → "SCHEDULE" below

"PHONE" → below, "CALC" → below, "SCHEDULE" below will
be initialized at power-up regardless of upgrades.

※ If "PHONE" has the optional CF-0453 bandpass & monitor function added,
※ If "SCHEDULE" has the optional CF-0451 tracking function added, it can be enabled. If no options are added, it cannot be enabled.

Other panel keys TRIG, AVG START, START, COMPARE, SELECT will be OFF at every startup.

3. Procedure for loading panel conditions.

Soft key settings

Turn MAIN → MEMORY → PANEL COND → RELAY LOAD​ ​ON.

Panel Condition Load Procedure

"MAIN" → "MEMORY" → "PANEL COND" → "COPY" to set Source to USB and Target to INT.

Press the soft key "ALL ON" once.
Press "COPY EXECUTE" to execute the copy.

4. The following items need to be reset (checklist)

□ LAN settings

「MAIN」→「CONDITION」→「LAN SETTING」

IP ADDRESS192.168.3.59
SUBNET MASK 255.255.255.0

□ RS-232C settings

「MAIN」→「CONDITION」→「RS232C SETTING」

BPS9600
CHARACTER BIT 8BIT
PARITY CHECKNONE
STOP BIT1
FLOW CONTROLNONE
TERMINATORCR

□ SAVE TYPE (Type of data to be saved)

「MAIN」→「MEMORY」→「DATA」→「SAVE TYPE」

TEXTOFF
BMPOFF


□ PROTECT (Prevention of Memory Overwriting)

「MAIN」→「MEMORY」→「PROTECT」

ON/OFFOFF


□ START TYPE (State at the time of main unit startup)

「MAIN」→「MEMORY」→「PANEL COND」→「START TYPE」

LAST TIME
NEW NEW
SELECT

Please reconfigure the start entry settings in the case of a SELECT operation.

□ RELAY LOAD (Reproducing the relay settings when reading the panel condition memory)

「MAIN」→「MEMORY」→「PANEL COND」→「RELAY LOAD」

ON/OFFOFF

* The red text indicates the factory default settings.

Last updated: 2015-05-28