CF-4500 Frequently Asked Questions
Question number FAQ-0282
What are the timing and duration of the "PANEL COND READ" output in OUTPUT RELAY?
It is shorted approximately 1.2 seconds before the panel condition loading is complete.
It is shorted for 1.2 seconds. After it is opened, it will accept INPUT RELAY and RS232C commands.
It is shorted for 1.2 seconds. After it is opened, it will accept INPUT RELAY and RS232C commands.
Last updated: 2012-06-12