CF-3000 Frequently Asked Questions
Question number FAQ-0182
I saved the trend data measured with 'A-Waighting for overall' set to 'on'. The replayed text data is not in 'A-wait' format.
The reason is
that the measured original data is FLAT characteristic data. From this original data, A-Wait is calculated and displayed as A characteristic data.
A-Waiting for overall is similarly calculated.
As a response,
(1) When inputting the ACout of the sound level meter,
set the filter characteristics of the sound level meter to A characteristics. Since ACout is the output through the A-Waighting Filter, using FLAT for the Filter setting of the CF3000 series will result in A characteristic analysis.
(2) When an MI Series microphone is connected directly
Inputting it into the CF3000 Series through a microphone amplifier with SR-2200 frequency correction, etc., produces the same result as (1).
Last updated: April 9, 2001