Skip to main content

Select your region & language

Global

Region

Digital Gauge Frequently Asked Questions

Question number FAQ-0101

What is the successor model to DG-410?

The models with a comparison function are DG4280 and DG4240. The model that has the same comparator output function as the DG410 is the DG4280.

(1) DG4280

Output conditions (judgment criteria)Judgment Output

Setting polarity = measurement polarity
and
| set value | ≦ | measured value |

The preset indicator light is lit in red.

Preset Output Relay ON

Turn ON preset output from BCD out connector.

(2) DG4240

The specifications for the contact output are divided into three categories: LOWER, OK, and UPPER.

criteria for judgmentJudgementDisplay and output of judgment results
LOWER setting value ≥ display valueLOWER・The COMPARATOR indicator light for LOWER is lit in red
・LOWER output relay (NO.10, 11) is ON
・SCD OUT connector LOWER output (pin 22) is ON
LOWER setting value < display value < UPPER setting valueOK・The OK indicator light of the COMPARATOR is lit in green
・OK output relay (No.12 13) is ON
・SCD OUT connector OK output (pin 22) is ON
UPPER setting value ≤ display valueUPPER・The COMPARATOR indicator light for UPPER is lit in red
・UPPER output relay (No. 14 15) is ON
・BCD OUT connector UPPER output (pin 23) is ON

For details, please refer to the FAQs for DG4280 and DG4240.

Last updated: 2001-01-12