Question number FAQ-0056

How to read data from a PLC and the data transfer time.

Please read the measurement values in the following order.

  • PLC → Hold the measurement value.
  • PLC ← DG-4160 holds the measurement value and returns it.
    At this time, the maximum time required from when the DG-4160 receives the hold bit set to when it sets the hold complete bit and the measurement value is approximately 70ms.

If the above steps are not followed, the DG-4160 will be set to normal mode, continuously outputting the latest display data. The data output interval will be 70ms.