Fuel Flow Indicator Frequently Asked Questions
Question number FAQ-0018
How do I perform the initial setup when using the FP-2240HA flow detector with the FM-2500 flow indicator?
When using the FM-series for the first time, you must configure the following items on the FP-flow detector during initial setup.
Similarly, if you change the FP-flow detector you are using, you will need to configure it in the same way.
Initial setup method when using the FP-2240HA flow detector with the FM-2500 flow indicator.
<Caution>
- This is how to set up the standard specifications. Please note that the settings will differ for custom specifications.
- This is how to configure the settings from the factory default.
- This setting measures instantaneous flow rate (volume). For settings for other measurements, please refer to the instruction manual.
- The display refresh time is 1 second (factory default). Please refer to the instruction manual for instructions on how to change this setting.
(1) Set the date. (※This step is unnecessary if the date is correct) →Click here for instructions on how to set the date.
(2) Set the following in order: ① detector, ② factor value, ③ encoder pulse, and ④ pressure sensor settings.
<Operating Procedure>
1.
Press (MENU) to display the settings screen.
![]() |
>>Please select the slot you wish to use. If using multiple devices, each one requires separate configuration.
|
2. The DF-0400 Setup menu will be displayed.
![]() |
|
3. The Sensor setup menu will be displayed.
![]() |
If FP224X is displayed (configured), no further configuration is required. Please proceed to step 6. If something other than FP224X is displayed (set), please configure it using the following steps.
|
4. Display the list of detectors to configure.
![]() |
>> Select the detector to use
|
5. Specify the detector to use.
![]() |
(The detector used is the FP-2240HA.)
|
6. Set the factor value.
![]() |
>>Set the factor value as indicated on the detector's nameplate or inspection report. |
![]() |
|
[Example of a 4-digit factor value]
If the factor value is 4 digits, set the most significant digit to 0: 09824 (Example setting)
7. Factor Value Determination
![]() |
8. Encoder pulse settings
![]() |
If the encoder pulse is displayed (set) as 120*10, no further settings are required. If the setting (display) is anything other than 120*10,
|
9. Select from the list of options.
![]() |
|
The displayed decimal places will start from 0.00.
*If you want to reduce the number of digits by one, please select 120.

10. Pressure sensor settings
![]() |
|
![]() |
11.
Pressing this button will bring up the measurement display screen.