Frequently Asked Questions about Fuel Flow Meters
Question number FAQ-0018
What is the initial setup method for using the FP-2240HA flow detector with the FM-2500 flow indicator?
When using the FM series for the first time, it is essential to perform the initial settings for the following items of the FP flow detector.
Similarly, if you change and use the FP flow detector, the settings are also required.
Initial setup method for using the FP-2240HA flow detector with the FM-2500 flow indicator.
Notice
- This is the method for setting standard specifications. Please note that the settings differ for custom specifications.
- This is the setting method from factory shipment.
- This is a setting for measuring instantaneous flow (volume). For settings related to other measurements, please refer to the instruction manual.
- The display update time is set to 1 second (factory default). For changes, please refer to the instruction manual.
(1) Set the date. (※Not necessary if the date is correct) →Click here for the setting method
(2) ① Set the detector, ② factor value, ③ encoder pulse, ④ pressure sensor settings in order.
Operating Instructions
1.
Press (MENU) to display the settings screen.
![]() |
Please select the slot to use. If using multiple, each requires configuration.
|
The 2.DF-0400 Setup menu will be displayed.
![]() |
|
3. The sensor setup menu will be displayed.
![]() |
If FP224X is displayed (set), no configuration is necessary. Please proceed to step 6. If something other than FP224X is displayed (set), please configure it using the steps below.
|
4. Display the list of detectors to be set.
![]() |
>> Select the detector to use
|
5. Specify the detector to be used.
![]() |
(The detector used is FP-2240HA.)
|
Set the factor value.
![]() |
Set the factor values as indicated on the detector's nameplate or in the inspection report. |
![]() |
|
[Example of a 4-digit factor value]
If the factor value is 4 digits, set the highest digit to 0: 09824 (example setting)
7. Determination of factor values.
![]() |
Setting of Encoder Pulses
![]() |
If the encoder pulse is displayed (set) as 120*10, no settings are required. If it is set (displayed) to something other than 120*10,
|
Select from the list.
![]() |
|
The number of decimal places will be 0.00 to.
*If you want to reduce the number of digits by 1, please select 120.

10. Setting of the pressure sensor
![]() |
|
![]() |
11.
Pressing it will bring up the measurement display screen.
Last updated: June 18, 2007