High-Performance Sound Level Meter
LA-7000シリーズ

The LA-7000 series sound level meter features a large color LCD touch panel, significantly improving operation and allowing for easy and intuitive measurement.
The LA-7000 series not only measures and calculates noise levels, but also allows you to "measure while listening to sound" during the measurement process. By listening to the sound, you can check the condition of the object being measured, locate the sound source, and confirm that the recording was successful. This enables safe and reliable measurements in measurement environments where failure is not an option.
Furthermore, by equipping optional features, it evolves beyond just a sound level meter (noise meter) into a sound analyzer, frequency analyzer, and sound recorder*. (*Optional)
The LA-7000 series sound level meter allows for noise measurement, sound recording, frequency analysis, and source detection of abnormal noises, significantly improving the efficiency of measurement work.
Furthermore, by combining the data saved with the LA-7000 series with our diverse range of analysis software, it becomes possible to perform waveform analysis, sound quality evaluation, fluctuating sound analysis, and sound frequency simulation.

Features

Easy to hold!

Hand strap included

This prevents the device from falling during measurement.

Size down

It's 35% smaller than the previous model. The design prioritizes ease of handling, making it comfortable even for those with smaller hands.

Easy to use!

Achieves long-term measurement

It can run for approximately 12 hours with alkaline batteries, and can also be powered via USB.

Intuitive operation via touch panel

You can change the calculation mode, range, measurement time, etc. by tapping the display.

Home key included

Pressing the home key will return you to the main screen.

Key lock function

You can lock the keys by long-pressing the home key.

Windscreen correction function

When using a windscreen for measurement, the effects of its installation can be compensated for.

Japanese language display

You can switch between Japanese and English display.

A convenient capture function for taking screenshots.

You can output the captured image.

External power ON/OFF function

If the device is integrated and needs to be turned ON/OFF in conjunction with an external power supply, this can be done by switching a switch.

Easy to understand!

4.3-inch color LCD

Overlaying text can also be displayed clearly. Range overruns are warned in red.

Recording starts with just one tap.

Calculation data and recording will begin with a single tap (for automatic memory saving).

Equipped with listening function

By pointing the microphone (sound level meter) towards the direction where the sound is loudest, you can not only see the numerical value but also experience the sound firsthand.

Video

List of features

Standard Options

Standard Options

SINGLE mode, DUAL mode, QUAD mode display standard functions

By combining different frequency weighting and time weighting characteristics, you can simultaneously display two (DUAL) or four (QUAD) calculated values.
This is useful for comparing measurement results of different weighted characteristics, such as A-weighting and Z-weighting values.

SINGLE mode MAIN
DUAL mode MAIN, SUB1
QUAD mode MAIN, SUB1, SUB2, SUB3
Example of QUAD mode display
Example of QUAD mode display

<Trend screen: Displays the trend waveform of MAIN or SUB. Only the selected value is shown. (MAIN or SUB)
*Trend display is based on a combination of instantaneous values or calculated values.

List > Screen: Displays all calculated values (L p, L eq, L E, L max, L min, L peak, L N) during measurement for the selected MAIN or SUB.
*To select, tap on the MAIN or SUB display; it will change to bold (indicating selection).

Listening function (phone output)- Standard feature

Measurement is possible while listening to sound.
  • Effective for measurements in anechoic chambers and remote locations.

  • Effective for monitoring remote locations due to environmental noise, etc.

  • *Extension cable and headphones are sold separately.
    *Extension cable and headphones are sold separately.

1/1 Real-time octave analysis function (standard feature)

RTA1/1 mode
RTA1/1 mode

This is one of the frequency analysis modes, displaying the data divided into octave bands. The vertical axis represents sound pressure level (loudness), and the horizontal axis represents frequency (pitch). You can see at a glance which pitches are loudest. You can also overlay and compare the memory data of RTA1/1 with the MAIN data.

Filter1/1 mode
Filter1/1 mode

By listening to sounds filtered through an octave band filter, you can isolate and hear only the sounds that are bothering you, even in noisy environments.
It helps to pinpoint the source of unusual noises by holding the microphone up to them.
Additionally, you can compare the sound pressure levels of Allpass (entire frequency range) and Bandpass (specific band).

Applicable standards IEC 61260-1:2014 Class 1 / JIS C 1513-1:2020 Class 1
Analysis band 16 Hz to 16 kHz (11 bands), Allpass1, Allpass2
*When the ultra-low frequency function is enabled, 1 Hz to 16 kHz (15 bands) [list only], Allpass1, Allpass2
Calculated item Filter Mode (Bandpass, Allpass switching) ・・・L p, L eq, L E, L max, L min, L peak, L N list
RTA Mode... L p, L eq, L E, L max, L min (Main)
Measurement range Normal range (Wide range is not supported)
Memory mode Manual, Auto, Logging, Record* (*When LA-0704 is installed)
Overlay function RTA mode *

The octave function includes an overlay function, which allows you to overwrite RTA memory data and measurement data (raw data).

Ambient noise compensation function RTA mode *

This can be used when you want to correct for ambient noise at a site during noise measurement.

What is background noise correction?
The sound pressure level measured while the equipment being measured is operating is a combined value of the sound emitted from the equipment and the background noise. Subtracting the effect of the background noise from this measurement to calculate the radiated sound pressure level of the equipment is called background noise correction.

Ambient noise correction function RTA mode

Points to note
When measuring ambient noise levels, it is important to obtain a sufficiently stable value.
In the frequency band, if the ambient noise data is larger than the signal, correction cannot be performed and the data will not be displayed.

*The overlay function and ambient noise reduction function can also be used in RTA1/3.

Convenient features

Calibration history function

Calibration is considered crucial when verifying the performance of a sound level meter. Acoustic calibration (verification using an acoustic calibrator) is performed before measurement. Having a calibration history makes management much easier in this case.

Convenient feature 1
Convenient feature 1

The recommended calibrator to be used can be registered in the sound level meter.
Registration items: Model of the calibrator used, calibration level value, serial number

Convenient Features 2
Convenient Features 2

Once calibration mode is completed by selecting OK, the calibration data is automatically saved to the internal memory (up to approximately 100 times). This calibration history data can be exported to an SD card (in .CSV format).
Items saved: (Calibration value, adjustment VR position, acoustic calibrator/internal reference signal used, calibration date and time)
*You need to register and select the calibration device you will be using.

Option

1/3 Real-time octave analysis function LA-0702 option

It is possible to analyze and evaluate sounds in even more detail by dividing them into octave bands.
It is also possible to hear the sound filtered through a 1/3 octave band filter. This is the most commonly used analysis function and filter in audio.

RTA1/3 Mode

RTA1/3 Mode

This is one of the frequency analysis modes, displaying the data divided into octave bands. The vertical axis represents sound pressure level (loudness), and the horizontal axis represents frequency (pitch). You can see at a glance which pitches are loudest. You can also overlay and compare the memory data and MAIN data from RTA1/3.

Filter1/3 mode

Filter1/3 mode

By listening to sounds filtered through an octave band filter, you can isolate and hear only the sounds that are bothering you, even in noisy environments.
It helps to pinpoint the source of unusual noises by holding the microphone up to them.
Additionally, you can compare the sound pressure levels of Allpass (entire frequency range) and Bandpass (specific band).

Specification

Analysis band

12.5 Hz to 20 kHz (33 bands), Allpass1, Allpass2

*When the ultra-low frequency function is enabled, the range is 0.8 Hz to 16 kHz (45 bands) [list only], Allpass1, Allpass2

Applicable standards This is the same as the 1/1 octave mode, so please refer to that.
Calculated item
Measurement range
Memory mode

FFT analysis function LA-0703 option

FFT analysis allows you to check the loudness of each frequency component of a sound.
It's also possible to calculate average results and trigger the detection of individual sounds.

  • FFT analysis function LA-0703
  • Supports auto-memory
    → Enables automatic saving of numerical data.

  • Supports record memory (*LA-0704 required)
    →FFT analysis is possible while recording.

  • Improved resolution; now supports 800 lines and 1600 lines.

Specification

frequency range 1k, 2.5k, 5k, 12.5k, 25kHz
Number of lines (number of sampling points) 400(1024)、800(2048)、1600(4096)
trigger ON/OFF count (1-16 times) / Level (40.0-130 dB), Position (-64 fixed point)
Average mode SUM, MAXhold, EXP (index)
Measurement range Normal range (Wide range is not supported)
Compatible memory Manual, Auto, Logging, Record
Expand screen display ×1、×2、×4

*For the LA-7700, the trigger level is (30.0 to 120.0 dB).

Sound recording function LA-0704 option

Sound recording function LA-0704

Audio data can be recorded, and the recorded sound can be heard on the sound level meter itself. The recorded files can be analyzed offline using O-Solution.

Specification

Memory mode Record
Sampling Frequency 64 kHz
Recording time When using 16 bit (4 GB approximately 8 hours)
When using 24 bit (4 GB approximately 5.5 hours)
Recording range Normal Range
(Wide range with 16-bit setting is not supported)
File formats wav (audio data) and
CSV (Trend data for playback: Lz at 1 s intervals)
* Simultaneous analysis recording in RTA and FFT modes is also possible.

(Note) When reading with Oscope, the limit is 2 GB (LA recorded data).
Please record within the following time frame.
Recording time:
When using 16 bit (2 GB 4.5 hours)
When using 24 bit (2 GB 3 hours)

Offline analysis function LA-0704 option

In measurement situations, there are increasingly fewer opportunities to take your time. The offline analysis function allows you to record sound and analyze it later.
The offline analysis function allows for "recalculation" and "re-analysis" (FFT, RTA, etc.) of Leq and other metrics. *Further detailed analysis is available through secondary processing software.

  • Offline analysis function LA-0704

Specification

  • SLM (Single/Dual/Quad mode)

  • RTA1/1, Filter1/1 mode

  • RTA1/3, Filter1/3 mode... LA-0702 required

  • FFT mode... LA-0703 required

*Note: The trigger function cannot be used.
*Note: Time setting is only available for MeasureTime.
* L-SIM (LA-0707) and SQE steady tone (LA-0708A) are supported in Ver4.0 and later.

Level detection function LA-0705 option

Output cable (MULTI I/O - BNC connector) 2 m included

  • If the set value is not exceeded (green) / *If the setting is OFF (gray)
    If the set value is not exceeded (green) / *If the setting is OFF (gray)
  • If the set value is exceeded (in red)
    If the set value is exceeded (in red)

Specification

targets for comparison SLM mode (SINGLE/DUAL/QUAD)
Main's Lp instantaneous value
Filter 1/1,1/3(BP:L p),RTA1/1,1/3(AP1:L p),FFT(OA:Inst)
(Calculated values are not supported)
Settings F 30.0 dB to 130.0 dB
DELAY setting OFF, 10ms, 100ms, 1s, 2s, 3s, 5s, 10s
HOLD setting 100ms, 200ms, 500ms, 1s, 5s, 10s, 30s, Manual
Output mode OFF, ON (Mode1LOW), ON (Mode2HI-z)

Level simulator function LA-0707 option

By lowering the level of the problematic frequency band and listening to the sound after the countermeasure (virtual sound), you can determine the overall value. This is useful for simulating the effect of the countermeasure in advance by comparing it to the sound before the countermeasure (original sound).

  • Level simulator function LA-0707 No.1
  • Level simulator function LA-0707 No.2
  • Level simulator function LA-0707 No.3

Specification

Measurement Mode L SIM
Switch display (Phone icon) UD (Simulation Possible) or AP (Original Sound)
Adjustment range -50 dB to +20 dB (1 dB steps)
Variable level compatible 16 Hz to 16 kHz (11 bands)
Display band 11-band, Overall, Allpass
Batch editing function 0 dB (for reset), -50 dB (total cut), display when UD is set.

 

MAIN Overall value (calculated value after increase or decrease)
List View Editing amount for each band, and the values after increase or decrease.
Output of UD settings Phone terminal
headphone output Overall (virtual sound) output
AC output Overall output (UD settings output)
AC/DC output Choose one from the following:
DC Output: Overall Output (Output for UD settings)
AC-Z output: Z output (Z characteristic of the raw sound)
Through output: Microphone output (output of raw sound converted to level).

Sound quality evaluation function (LA-0708A option)

Noise level evaluations can sometimes differ from human perception. In such cases, sound quality evaluations, such as loudness ratings, can be very effective.
Loudness is a useful tool for evaluating the perceived loudness of sound.
Sharpness is useful for identifying metallic, high-pitched, unpleasant sound components.

The CI value* is a value that takes into account time-averaged sound level and sharpness. It is simple and effective for applications such as ambient noise and machine noise (dental drills). (Further research is being conducted on its application scope.)
The LA-0708A's sound quality evaluation function is designed for stationary sounds. For non-stationary sound loudness evaluation, please use the sound quality evaluation software from the Oscope series.

  • Sound Quality Evaluation Function LA-0708A No.1
  • Sound Quality Evaluation Function LA-0708A No.2

Specification

Usage range Stationary sound: Selection between "Diffuse field (GD) and Free field (GF)"
List of operations
Loudness (compliant with ISO 532-1 stationary sound standards) “N”
Loudness level (same as above) “LN”
Sharpness (DIN45692 compliant) “S”
Sharpness (Aures) “SA”
Sharpness (Bismarck) “SB”
Comfort Index* “CI”
Simultaneous measurement Main Choose one from the above list of operations
SUB1/SUB2/SUB3 One value from each SLM (the frequency-weighted characteristic of SUB1 will be fixed at A).
Measurement range Normal Range

* This product cannot be attached to the LA-7200 (Class 2).
* CI (Comfort Index) is one of the sound evaluation metrics developed by Professor Sonoko Kuwano.* CI is the sum of the "sharpness value" and "1/10 of the A-weighted time-averaged sound level." There are correlation examples with environmental noise and mechanical noise, and it is considered an effective measure for evaluating unpleasant sounds. The scope of application is still being investigated, so please use it as a reference value.

Sonoko Kuwano* (Doctor of Engineering) Professor Emeritus, Osaka University. Former President of the Acoustical Society of Japan, Vice President of the Japan Society of Noise Control Engineering, and Vice President of the International Acoustical Society.

Infrasound measurement function LA-0709 option

It is also possible to measure extremely low-frequency sounds (1-20 Hz sounds) that are inaudible to the human ear.

Measurement frequency range 1 Hz to 20 kHz
(By switching the measurement mode, it is possible to measure between 10 Hz and 20 kHz or between 1 Hz and 20 kHz.)
Frequency weighting characteristics G-type characteristics can be selected.
Time-weighted properties 10 seconds selectable
Applicable standards ISO7196:1995 Acoustics - Frequency-weighting characteristic for infrasound measurements
Supported Modes SLM-Single, Dual, Quad, Filter1/1, RTA1/1, Filter1/3(LA-0702), RTA1/3(LA-0702), FFT(LA-0703)

*This unit can only be installed on the LA-7500. If you wish to install it on an existing LA-7500, we will arrange for pickup.

Option price

Model name Product name Price (tax included)
LA-0702 1/3 Real-time Octave Analysis Function  
LA-0703 FFT analysis function  
LA-0704 Sound recording function  
LA-0705 Level determination function  
LA-0707 Level simulator  
LA-0708A Sound quality evaluation function  
LA-0709 Infrasound measurement function  

Interface

↓ There is a link to more details.

Interface

Specification

  LA-7700 Class 1 LA-7500 Class 1 LA-7200 Class 2
Measurement Law Type Approval Number *1 No. TF2210
No. TF228
No. TF229
No. TF227
No. TS225
No. TS224
Applicable standards JIS C 1509-1:2017 Class 1
JIS C 1516:2020 Class 1
JIS C 1509-1:2017 Class 2
JIS C 1516:2020 Class 2
IEC 61672-1:2013 Class 1 IEC 61672-1:2013 Class 2
ANSI S1.4-2014/Part 1 Class 1 ANSI S1.4-2014/Part 1 Class 2
Measurement frequency range 10 Hz~20 kHz 10 Hz to 20 kHz
1Hz to 20kHz (when ultra-low frequency function is enabled *2)
10 Hz~8 kHz
Measurement level range (JIS, IEC) A: 20–128 dB
C: 28–128 dB
Z: 34–128 dB
A: 24–138 dB
C: 32–138 dB
Z: 38–138 dB
Z: 50–138 dB (when ultra-low frequency function is enabled)
G: 35–138 dB (when ultra-low frequency function is enabled)
A: 23–138 dB
C: 30–138 dB
Z: 36–138 dB
Self-noise level A: 12 dB or less
C: 20 dB or less
Z: 26 dB or less
A: 16 dB or less
C: 24 dB or less
Z: 30 dB or less
G: 27dB or less (when ultra-low frequency function is enabled)
A: 17 dB or less
C: 24 dB or less
Z: 30 dB or less
Microphone MI-1281 MI-1271 MI-1471
Microphone preamplifier MI-3270
Linearity range Wide range: 110 dB / Normal range: 80 dB
Level range 10~120 dB(wide)/50~120 dB/40~110 dB/
30-100 dB/20-90 dB/10-80 dB/0-70 dB
20-130 dB(wide)/60-130 dB/50-120 dB/
40-110 dB/30-100 dB/20-90 dB/10-80 dB
Reference range 50–120 dB
Time-weighted properties Example: LAF F (fast), S (slow), I (impulse), 10 ms F (Fast), S (Slow), I (Impulse), 10 ms, 10 s (When ultra-low frequency function is enabled) F (fast), S (slow), I (impulse), 10 ms
Frequency weighting characteristics Example: LAF A, C, Z characteristics A, C, ZG characteristics (when ultra-low frequency function is enabled) A, C, Z characteristics
Measurement items L p, L eq, L E, L max, L min L peak,
L N (L 5, L 10, L 50, L 90, L 95, L high, L low, L ave. Any two LN values)
Sampling interval 15.6 μs (L p, L eq, L E, L max, L min, L peak), 100 ms (L N)
Measurement time Example: When measuring for 24 hours with repeated measurements at 10 minutes on the hour,
MT: 10 min, PT: 1 hour, TT: 24 hours
  Measurement time (Meas. Time) Manual (OFF), arbitrary setting 0.1 to 199 hours 59 min 59.9 sec Resolution: 0.1 sec
Period Time Resolution: 1 min to 24 hours
Total measurement time Resolution: 0.1 sec to 999 hours 59 min 59.9 sec
Start type (start mode) Manual start, timer start, countdown start, level start
Display function Display unit 4.3-inch LCD display with color backlight (touch panel compatible)
  Digital display 4 digits/resolution 0.1 dB/display cycle 1 s
Bar indicator Wide range: Display range 100 dB / Normal range: Display range 70 dB
Battery level indicator 4-level display
Operate Mode (Online) Mode

Standard: SLM-Single, Dual, Quad, Filter1/1, RTA1/1

Options: Filter1/3 (LA-0702), RTA1/3 (LA-0702), FFT (LA-0703), LSIM (LA-0707), SQE Stationary Tone (LA-0708A)

Standard: SLM-Single, Dual, Quad, Filter1/1, RTA1/1

Options: Filter1/3 (LA-0702), RTA1/3 (LA-0702), FFT (LA-0703), LSIM (LA-0707)
* The "SQE steady-state sound (LA-0708A)" option is not available.

Offline mode
(LA-0704 required)

Standard: SLM-Single, Dual, Quad, Filter1/1, RTA1/1
Options: Filter1/3 (LA-0702), RTA1/3 (LA-0702), FFT (LA-0703)
* LSIM (LA-0707) and SQE steady tone (LA-0708A) are supported in Ver4.0 and later.

Overlay display
(Overlay)
Overlay display Standard: RTA1/1
Option: RTA1/3 (LA-0702)
Ambient noise correction display Standard: RTA1/1
Option: RTA1/3 (LA-0702)
Memory function Saves to an SD/SDHC card (SDHC cards up to 32 GB can be used).
  Memory mode MANUAL (CSV file), AUTO (instantaneous value, calculated value, CSV file) ... Standard function
LOGGING (instantaneous value 10 ms or 100 ms, CSV file) ... Standard feature
RECORD (WAVE file: 64 kHz sampling) ... LA-0704 required
Panel condition memory Internal memory and SD/SDHC card
(Internal memory: 5 rewritable memory slots, 5 EZ condition slots, 1 condition slot when powered off)
(SD/SDHC: The number depends on the capacity of the SD/SDHC card)
Simple measurement mode 5 modes
(EZ1:L Aeq +L Cpeak, EZ2:Record, EZ3:Logging 100 ms, EZ4:NC, EZ5)
Clock function Built-in (Year/Month/Day/and Hour:Minute)
Continuous operation time: Approximately 1 year (Charging time: 24 hours from a completely discharged state)
Calibration history function Internal memory (saves approximately 100 items)
Contents (calibration value, adjustment VR position, acoustic calibrator/internal reference signal used, calibration date and time)
Resume function Save measurement conditions to internal memory
Calibration Reference signal
(Output for when connecting external devices)
Electrical calibration using an internal oscillator (1 kHz sine wave)
Normal range: -6 dB from full scale, wide range: -16 dB from full scale
Recommended Calibrator SC-2600, SC-2500A SC-2600, SC-2500A, SC-2120A
Output/Input Phone output
(Headphone output)
Actual sound, or recorded sound (playback sound) 1/1 octave filter (standard function),
When using the 1/3 octave filter mode (optional: LA-0702),
The actual sound or recorded sound (playback sound) of the selected band.
When "UD" is selected in the level simulator (LSIM: LA-0707), the level editing sound is...
When "AP" is selected, the actual sound
Maximum output: 10 mW (63 Ω: at 1 kHz), Connector: Stereo ⌀3.5
AC output Output one of A/C/Z in conjunction with the screen (Main).
  AC output level Output level: 0.707 Vrms ± 5% (normal range), 2.236 Vrms ± 5% (wide range)
Range full-scale input, 1 MΩ load
Distortion (full range scale): 0.2% or less.
Load resistance: 10 kΩ or more,
Offset voltage: within ±30 mV,
Output impedance: 50 Ω ± 2%
AC/DC output DC, AC-Z, Through switching
  DC output level 2.5 V ± 20 mV (normal range, wide range) at full-scale input, 1 MΩ load.
Scale factor: 0.25 V ± 10 mV/10 dB
Load resistance: 10 kΩ or more,
Output impedance: 50 Ω ± 2%
AC-Z output level Output level: 0.707 Vrms 5% (normal range), 2.236 Vrms ±5% (wide range)
With a full-scale input range and a 1 MΩ load,
Distortion (full range scale): 0.2% or less.
Load resistance: 10 kΩ or more,
Offset voltage: within ±30 mV
Output impedance: 50 Ω ± 2%
Through extension level

0.707 Vrms ± 5% (normal range, wide range) at full-scale input, 1 MΩ load.
Distortion: within 0.2% output

External control input
operation Reset and start
Control voltage : Voltage-free contact input
Input pulse width : 200 ms or more
Absolute Maximum Input Voltage : 24.0 V
Level detection output (LA-0705) Open collector (Output voltage rating: DC +24 V or less, sink current: 20 mA or less)
"Output cable (MULTI I/O - BNC connector 2 m) included"
  Level determination comparison targets SLM mode (Main: L p), Filter 1/1,1/3 (Main(BP): L p), RTA1/1,1/3 (AP1: L p), FFT (OVERALL: Instant)
Interface RS-232C Transfer speed: 9600, 115200 bps, MULTI I/O cable (sold separately)
USB
Ver2.0 : USB High Speed compliant
USB connection cable USB (A) male to Micro USB (B) male (sold separately)
Function : Command control, or mass storage *3
External memory SD/SDHC memory card, maximum capacity SDHC 32 GB *4
Compatible extension cable
(Microphone extension) *5
AG-3400 series (CE compliant up to 30 m), maximum extension: 103 m (AG-3305),
Cable extensions exceeding 5 meters have a compensation function.
Power supply 4 x AA batteries (alkaline or Ni-MH rechargeable batteries)
• USB bus powered (operating input voltage range: DC 4.75 to 5.25 V)
• AC adapter (PB-7090, power consumption at AC100V, approximately 7 VA)
External power supply operation function This function controls the power of the unit in conjunction with the power supply from the AC adapter. (The external power-on mode does not work with batteries.) This can be switched using a switch inside the battery compartment (standard function).
Windproof screen correction function This function compensates for the effects of the windproof screen "Windproof Screen (⌀70) / All-Weather Windproof Screen (LA-0207A)". (LA-7700 is not compatible with the All-Weather Windproof Screen.)
Battery life (continuous use) *6 Alkaline batteries (LR6): Approximately 12 hours; Ni-MH rechargeable batteries: Approximately 12 hours
Operating (storage) temperature range −10 to 50 ºC (−20 to 60 ºC)
Humidity range for use (storage) 20% to 90% RH (10% to 90% RH), provided there is no condensation.
Size Approximately 90 mm (W) x 279 mm (H) x 42 mm (D)
Weight Approximately 540 g (including batteries)
Accessories AC adapter (PB-7090) *7
Signal cable (AX-501)
Windbreak screen (⌀70 mm)
Hand strap, 4 AA alkaline batteries
Carrying case (including shoulder strap)
SDHC memory card (16GB)
Instruction Manual (CD version)
Setup Guide
Price (tax included)
 
 
 
Inspection fee at time of shipment *8
 
 
 

*1 Measurement Law (enforced in 2021)

*2 LA-0709 Supported by the ultra-low frequency sound measurement function

*3 Supported in firmware Ver. 2.0 and later.

*4 When using the SD memory function, please use the recommended SD card.

*5 The cable extension distances listed here are based on the use of our designated dedicated cables.

*6 Operating time varies depending on the operation mode, memory mode, and backlight settings.

*7 Please inquire separately regarding overseas-specification power cords.

*8 As a general rule, our company replaces the inspection conducted at the time of shipment with an "Approval Mark" obtained through our own inspection.
After shipment, we inspect and adjust our products before requesting them to be inspected by the Japan Quality Assurance Organization (JQA) (Standard sound level meter: ¥15,300 (tax-free) + ¥22,900 (excluding tax); Precision sound level meter: ¥23,000 (tax-free) + ¥35,400 (excluding tax)). Since we offer products without inspection as standard, customers who require products with inspection (certificate of conformity) or who wish to have their products inspected by a public institution from the time of shipment should inform us when placing their order.

Accessories

  • Accessory No. 1
  • Accessory No. 2

Related Products

Sound calibrator (SC-2600 / SC-2500A / SC-2120A)

  • Sound calibrator (SC-2600 / SC-2500A / SC-2120A)

The LA-7200 (Class 2) includes the SC-2120A, SC-2500A, and SC-2600.
For LA-7500 (Class 1) / LA-7700 (Class 1), please use SC-2500A or SC-2600.

LA-0207A All-weather windproof screen

LA-0207A All-weather windproof screen

Using the all-weather windproof screen reduces wind noise during outdoor noise measurements and prevents microphone damage from rain, snow, etc. Please set the extension cable compensation function on the LA-7000 unit before use.

Screen diameter: ⌀ 200 mm
Mounting screw thread: 1/4-20 UNS depth 6 (camera tripod screw)

* With LA-0207A installed, it conforms to JIS C 1509-1:2017/IEC 61672-1:2013.
* Consumable parts (windscreen, waterproof mesh) will need to be replaced periodically.
* This product cannot be attached to the LA-7700 (Class 1).

AG-3400 Series Microphone Extension Cable

  • AG-3400 Series Microphone Extension Cable
    MI-0301 Holder included

The microphone unit and the main unit can be separated, and the microphone can be extended over long distances using a dedicated microphone extension cable.
* Please use the LA-7000 after setting the extension cable correction function.

  • AG-3401 Microphone Extension Cable (5 m):

     

  • AG-3402 Microphone Extension Cable (10 m):

     

  • AG-3403 Microphone Extension Cable (20 m):

     

  • AG-3404 Microphone Extension Cable (30 m):

     

Recommended USB cable (item code PE19B2458*)

Recommended USB cable (item code PE19B2458*)

U2C-AMBF2U12BK (manufactured by ELECOM)
Micro-USB (A-microB) cable, 2A output compatible (1.2 m), with ferrite core.
USB (A) to USB (microB) (Sound level meter side)
*Our internal code

MULTI I/O Cable (Item Code PE3534088)

MULTI I/O Cable (Item Code PE3534088*)

Used for RS-232C interfaces. (2 m)
Serial port side (D-sub 9-pin) MULTI/IO (Sound level meter side)

*Our internal code

Sound and Vibration Analysis System O-Solution/DS-5000

Detailed analysis using analysis mode

Detailed analysis using analysis mode

By importing data (in WAV format) recorded with the LA-7000 series (+LA-0704) into O-Solution, you can play back the sound, perform frequency analysis, and octave analysis on your PC.
Additionally, by using an IIR filter (optional), it is possible to listen to the filtered sound while playing back the recorded audio.

As a signal input source for the DS-5000

As a signal input source for the DS-5000

Instead of using a microphone, you can input the analog output of the LA-7000 series into the DS-5000 to perform real-time analysis. The LA-7000 series can simultaneously output two signals: one with an acoustic filter (such as A-weighting) and one with a flat (Z-weighting) signal. These can then be analyzed simultaneously by the DS-5000.

Time Series Data Analysis O-Solution

Sound quality evaluation function OS-0525

Sound quality evaluation function OS-0525

The OS-0525 sound quality evaluation function allows you to obtain sound quality evaluation indicators that take into account human auditory characteristics and correspond to various perceptions of sound, such as "loudness," "sharpness," "roughness," "variability intensity," and "tonality."

Sound fluctuation analysis function OS-0526

Sound fluctuation analysis function OS-0526

The OS-0526 variable sound analysis function can extract only the components with large time variations that are not affected by level, enabling the quantification of various time-varying characteristics that were difficult to detect using roughness or variation intensity. It also includes a function to compare sounds with and without variations. (Variable Sound Simulator)

Time-frequency analysis function OS-0527

Time-frequency analysis function OS-0527

It can evaluate transient phenomena that are difficult to capture with FFT analysis. It can clearly display the time changes of frequency components while maintaining the frequency resolution of the characteristic quantities (frequency). It is equipped with short-time Fourier transform and wavelet transform.

The link between subjective and physical quantities
The link between subjective and physical quantities

Sound One: A sound platform that connects sensibility and physics

This web application allows for sound recording, sound editing, data management, audio testing, and linking subjective and physical quantities. It is useful for evaluating abnormal sounds and sound design.

You can stream the sound you want to evaluate via the web, collect listener feedback, and check its correlation with parameters such as sound pressure level. This service is available on your smartphone, computer, or any other device you own.

Predicting various noise sources in the same field
Predicting various noise sources in the same field

SoundPLAN noise environmental prediction software

ASJ CN-Model 2007
SJ RTN-Model 2013

This is acoustic simulation software that predicts and calculates how noise generated from roads, railways, factories, etc., propagates.
Even in areas where various noise sources such as road noise, railway noise, factory noise, large retail stores, construction work, and power substations coexist, complex calculations can be performed without switching software.

Sound source visualization system BF-3200 / MI-5420A

  • Sound source visualization system BF-3200 / MI-5420A
  • 3200 BF Monitor
    3200 BF Monitor

Building upon the LA-7000 series' ability to pinpoint the location of unwanted noises (sounds filtered through an octave bandpass filter), this device takes things a step further. It allows you to "visualize" unwanted noises (frequency of a 1/3 octave bandpass filter).

Calibration

To guarantee the accuracy of the values measured by a measuring instrument, the instrument must be calibrated at appropriate intervals. Only by using a calibrated measuring instrument can the reliability and stability of the obtained measurements be improved. There are three calibration services available for high-performance sound level meters: "ASNITE calibration," "JCSS calibration," and "general calibration."