本文にスキップする

Select your region & language

Global

Region

Digital gauge counter
DG-4320/DG-4340

DG-4320
DG-4320
DG-4340 (with comparator)
DG-4340 (with comparator)

The digital gauge counter DG-4320/DG-4340 inputs signals from a linear gauge sensor and precisely measures and displays dimensions and displacements. It adds new features necessary for measurement, such as calculation functions and condition storage functions, demonstrating its capabilities for inline measurement in production sites.

* The display example of DG-4340 (in red) is when the comparator indicates NG.
When the OK judgment is made, it will display in green, similar to DG-4320.

Features

  • Ono Sokki Compatible with all linear gauge sensors (except GS-3800B series)

  • Step multiplication switching, panel condition memory, and various functions such as MAX, MIN, and RANGE calculations.

  • Compatible with 12 V power supply sensors (optional)
    Counts signals from single-phase pulse signals such as rotation detectors or from encoders.

  • With the 3ch comparator function, it is possible to determine the pass or fail of parts and products (DG-4340)
    When the comparator operates, the backlight of the display screen changes to red, making it easy to understand.

System Configurations

  • System Configurations

Connection and Communication

The BCD output of DG-4320/DG-4340 is generally equivalent to the pin assignment and connector of DG-4140.
However, due to new features, pins that were previously unused in older models are now utilized. Additionally, the comparator output terminals present in DG-4240/DG-4280 are not available.

Pin signal content Pin signal content
DG-4120 DG-4140 DG-4240 DG-4280 DG-4320/DG-4340
1 1×10 0 BCD output 19 4×104
2 2×100 20 8×104
3 4×100 21 N.C Start Input Lower output Preset1 Start Input
4 8×100 22 N.C Stop input OK output Preset2 Stop input
5 1×10 1 BCD output 23 N.C N.C Upper Output Preset3 Display Switch Input 1
6 2×101 24 N.C N.C N.C Preset4 Display Switch Input 2
7 4×101 25 Polarity Output +
8 8×101 26 Negative Polarity Output
9 1×10 2 BCD output 27 D.P 0.00 Decimal Point 1
10 2×102 28 D.P 0.000 Decimal Point 2
11 4×102 29 N.C 1×10 5 BCD output
12 8×102 30 error output
13 1×10 3 BCD output 31 Hold Input
14 2×103 32 Reset Input
15 4×103 33 busy input
16 8×103 34 N.C Comparator gate input Comparator gate input N.C Comparator gate input
17 1×10 4 BCD output 35 Print command output
18 2×104 36 common

Case Studies

Pass/Fail Judgment by 3ch OK/NG Comparator Function

By detecting the height of the screw, you can check the tightness of the screw.
Equipped with a 3-channel OK/NG comparator, it can accommodate three types of screws of different sizes.

  • Pass/Fail Judgment by 3ch OK/NG Comparator Function

Pass/Fail Judgment by 3ch OK/NG Comparator Function

In the flatness inspection of parts, a sensor is applied to the subject and rotated once.
By looking at the RANGE value (MAX-MIN), flatness measurement can be easily performed.

  • Pass/Fail Judgment by 3ch OK/NG Comparator Function

Length measurement using pulse counter function

The length of paper or steel plates is measured using a rotary encoder attached to the feed rollers.
By using the multiplication switching function and factor setting function, the pulse output from the encoder can be converted into length.

  • Length measurement using pulse counter function

Automatic identification by BCD output function

We will determine whether the dimensions of the processed parts are within the standards. By inputting BCD output into the sequencer, sorting can be performed based on the measurement values.

  • Automatic identification by BCD output function

Specifications

Model name DG-4320 DG-4340
method of representation LCD display, segment and dot matrix composite display, 2-color backlight, 6-digit display.
display range 0 to ±199999 (when 0.5 µm resolution is selected, the first digit will display only 0 or 5)
compatible sensor Our linear gauge sensors (excluding GS-3800 series) *1
2-phase and 1-phase rectangular wave voltage output type sensors (such as encoders)
Power output for sensors: DC 5 V ±10 %, 200 mA
Resolution switch Choose from 0.5 µm, 1 µm, 10 µm *2
input signal 2-phase phase-shifted rectangular wave, 1-phase rectangular wave (supported by internal setting switch)
Input frequency range DC ~300 kHz
Display numeric type Instantaneous value, maximum value, minimum value, range value (maximum value - minimum value)
computational function Offset function, factor function, multiplication switch function, decimal point position change function
digital output BCD output, update time 10 ms, open collector output (positive and negative logic switchable)
Comparator ― 3-stage, semiconductor relay method
Switching between Upper and Lower comparators
and OK/NG comparators in settings
Backlight changes to red during operation
Panel Condition Memory Function Save the setting parameters inside the counter (4 conditions)
External Dimensions / Mass 72(W) × 72(H) × 114(D) mm / approximately 320 g
Power supply voltage AC 100~240 V
Operating temperature range 0 °C to +40 °C, approximately 6 VA (at AC 100 V)
Storage Temperature Range −10 ℃~+55 ℃
Price    
Option ・DG-0430 電源供給オプション(DC 12 V 150 mA)
・AA-8101 BCDケーブル 3 m(DA-4130 D/A変換器用)
・AA-8107 BCDケーブル 3 m(片側オープン)
・AX-2050N 電源ケーブル 3 m(AC100V用)

*1 Please contact us if you are using the DG-4300 series and BS-102 series (discontinued) together.
*2 If you select an encoder as the sensor type, the unit and decimal point will not be displayed.
(The decimal point can be displayed using the "Decimal Point Position Setting Function")

Recommended interface for BCD external signal input/output

Output Signal Interface Circuit

(BCD output, polarity output, decimal point output, error output, print command output)

output format open collector output
Withstand voltage Max. 24 V
maximum sink current Max. 30 mA
residual voltage Max. 0.5 V
  • Output Signal Interface Circuit No.1

(Start input, Hold input, Reset input, Busy input, Comparator gate input)

Input low level voltage 0 ~1.4 V
Input Hi-level voltage 3 to 5.25 V
input impedance 1 kΩ or more
  • Output Signal Interface Circuit No.2

最終更新日:2023/07/27