| Contact output |
1 Make contact output × 3 (COMP1/COMP2/COMP3) - Individual judgment conditions can be set. |
| Judgment conditions |
UPPER |
[TM-4100/4200/4400] 6-digit setting 【TM-4300】 7-digit setting • When UPPER ≤ Main measurement value, the relay is turned ON. |
| LOWER |
[TM-4100/4200/4400] 6-digit setting 【TM-4300】 7-digit setting • When LOWER > Main measurement value, relay ON |
| OK |
When all comparators set to UPPER or LOWER are OFF, the relay turns ON. |
| ERROR |
ERROR: Relay ON when an error other than communication occurs. |
| Contact operation mode |
Automatic recovery mode |
After the relay is turned ON, it will return to OFF when the judgment conditions are no longer met. (Hysteresis can be set in the judgment conditions UPPER and LOWER, from 0 to 20%) |
| Hold mode |
[TM-4100/4200/4300] Once the relay is turned ON, it will remain ON even if the conditions for judgment are no longer met. |
| Shot output mode |
If the relay ON condition is met, it will turn ON for the specified time and then return to OFF. (Adjustable time: 10 to 2000 ms, in 10 ms increments) |
| Output delay function |
[TM-4100/4200/4300] The relay turns ON if the set value is exceeded for a set period of time or longer (configurable time: 0 to 1000 ms, in 10 ms increments). |
| Reset function |
[TM-4100/4200/4300] When in hold mode, the relay is returned to OFF. |
| Maximum contact capacity |
DC:30 V/1 A |
| AC:250 V/1 A |
| Output update time |
Approximately 10ms |
| Output format |
terminal block |