Notice of a problem when analyzing with time weighting set to Impulse in octave analysis of the DS-2000 series

Thank you for your continued patronage of our products.
We have discovered a problem with the octave analysis function of the DS-2000 series that you recently purchased, specifically when the time weight is set to Impulse. We sincerely apologize for causing this issue and promise to prevent such a problem from recurring.
We sincerely apologize for any inconvenience this may cause to our customers using this product. We kindly request that you install the latest version of the software with the fix implemented. Please refer to the "Version Upgrade Instructions" at the bottom of this page for details.
We sincerely apologize for bothering you during this busy time, but we would appreciate your understanding and cooperation.
If you have any questions or concerns, please contact our customer service department listed below.

Causes and symptoms of the malfunction

When performing an analysis with the time weight set to Impulse, there was a bug in the processing of the remaining levels when switching from a falling time constant of 1500ms to a rising time constant of 35ms.
When performing an "offline analysis" of octave analysis and setting the time weight to Impulse, the values for each band and the Allpass may become larger than the correct values.

Target Products

Model name Software Version
DS-0223 Real-time Octave Analysis

DS-0223L Offline License Version
DS-0223V for 2 ch
DS-0223W for 4~8 ch
DS-0223X for 10~32 channels
Offline Analysis in the VersionJ.txt file on the installation CD*: Ver1.14.0.35Before
This number is1.14.0.35From then on Completed
Numbers prior to this date are not supported.

countermeasure

We have prepared a version that fixes the above bug.
Please download the latest updated software and perform an overwrite installation on the computer where your DS-2000 is installed.

Previous version Revised version
OfflineAnalysisi version prior to 1.14.0.35 → 1.14.0.35

How to upgrade

Clicking the "Go to User Login/Registration Page" button below will take you to the "User Registration & Download" page.
"DS-2000 Graduo Multi-Channel Data Station" on the displayed page

  • How to upgrade

First-time users should click [New Registration], and existing users should click [Special User Page Login] to proceed to the DS-2000 software download page, from which you can download the installer.

We apologize for the inconvenience caused by the login and registration processes.

Please fill in the information.

 

 

The information on this page is based on the information available at the time of publication.
Please note that if product sales are discontinued, organizational changes are made, or specifications are changed after publication, the information may differ from the latest information.

This page is created by an automatic translation system based on the information in the Japanese version. Details